12/15/18 Channels

ECG machine 

Main unit

External sampling box

0123

12/15/18 Channels

ECG machine specification

Model: e1200/e1200 plus/e1800/e1800 plus

  • Defibrillator-proof protection

  • Pacemaker detection

  • Large touch screen, easy operation

  • Supporting an accurate digital filter to decrease the polarization voltage and other interferences

  • Supporting ECG waves displaying with grid

  • Automatic baseline adjustment for optimal printing

  • Convenient operation of system setup and file management
    Advanced Might-green algorithm for automatic measurements and interpretation

  • Large internal memory which can be enlarged by USB flash disk & SD card

  • Barcode scanner support

  • LAN/Wi-Fi connections

  • VCG/TVCG/VLP/HRV/Pharmacological Study

  • Supporting DAT、SCP、PDF、XML、HL7、DICOM File format Measurement function and interpretation function

CONFIGURATION

Model

7inch LCD

10.1inch LCD

9 channel

12 channel

15 channel

18 channel

e1200

YES

YES

YES

NO

NO

NO

e1200 plus

YES

YES

YES

NO

NO

NO

YES

YES

YES

YES

YES

NO

e1800

e1800 plus

YES

YES

YES

YES

YES

NO

STANDARD CONFIGURATION

e1200/e1200 plus

Standard Configuration

Part Number

Power cord(European)

07.01.000001

Power adapter

08.01.000001

12-lead Patient Cable (European)

07.03.000014

Adult Chest electrodes (6pcs/set)

08.05.000005

Adult Limb electrodes (4pcs/set)

08.05.000006

ePCECG-12

sampling box

11.1007.0001

USB cable

07.03.100001

Thermal Recorder Paper

(210 mm×295 mm ×100
pages)

08.01.000002

e1800/e1800 plus

Standard Configuration

Part Number

Power Cord (European)

07.01.000001

Power adapter

08.01.000001

18-lead Patient Cable (European)

07.03.000016

Adult Chest electrodes

08.05.000005

Adult Limb electrodes

08.05.000006

ePCECG-18

sampling box

11.1007.0002

USB cable

07.03.100001

Thermal Recorder Paper

(210 mm×295 mm ×100
pages)

08.01.000002

PERFORMANCE SPECIFICATION

  • Anti-electric-shock type : Class I equipment

  • Category AP/APG equipment : 

      Not category AP/APG equipment 

  • Emergency care equipment : Not emergency care equipment 

  • Mode of operation: continuous 

  • Recorder: Thermal dot-matrix recorder 

  • Printing Density :
    8 dots per mm / 200 dots per inch (amplitude axes) 
    40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s) 

  • Recorder Paper: 
    Folded thermal paper: 
    210 mm×295 mm ×100 pages, 
    215 mm×280mm×100 pages (optional) 

  • Leads: 9/12/15/18 Leads

  • Gain: 2.5mm/mV、5mm/mV、10mm/mV、

        20mm/mV,10/5mm/mV, AGC±5%

  • Speed: 2.5mm/mV、5mm/mV、10mm/mV、20mm/mV,10/5mm/mV, AGC±5%

  • Sampling Frequency: 16KHZ

  • A/D Resolution: 24bits

  • Anti-electric-shock degree: CF type with defibrillation-proo

  • HR Range: 30~300BPM

  • Accuracy: 1BPM

  • Time Constant: ≥3.2s

  • Input Circuit Current: ≤0.01μA

  • Input Voltage Range: ≤±5mVp-p

  • Calibration Voltage: 1mV±2%

  • DC Offset Voltage: ±600mV

  • Input Impedance: ≥50MΩ(10HZ)

  • CMRR: ≥120dB

  • Frequency Response:  (0.01~300Hz) +0.4dB / -3.0dB

  • Noise: ≤12.5μV (P-P)

  • AC Filter: 50Hz/60Hz/Off

  • DFT Filter: 0.01HZ/0.05HZ/0.32Hz/0.67Hz

  • Defibrillation-proof recovery time: ≤5s

  • EMG Filter/Lowpass: 

        25Hz/35Hz/45Hz/75Hz/ 100Hz/150Hz/270Hz/300Hz

  • Pacemaker Detection: 

         Amplitude: ±750uV~±700mV

         Width: 50us~2.0ms

         Sampling Frequency: 16000HZ

  • Patient Leakage Current:

         NC: ≤10uA(AC)/ ≤10uA(DC)

         SFC: ≤50uA(AC)/ ≤50uA(DC)

  • Patient Auxiliary Current:

         NC: ≤10uA(AC)/ ≤10uA(DC)