12/15/18 Channels
ECG machine

Main unit

External sampling box

0123

12/15/18 Channels
ECG machine specification
Model: e1200/e1200 plus/e1800/e1800 plus
-
Defibrillator-proof protection
-
Pacemaker detection
-
Large touch screen, easy operation
-
Supporting an accurate digital filter to decrease the polarization voltage and other interferences
-
Supporting ECG waves displaying with grid
-
Automatic baseline adjustment for optimal printing
-
Convenient operation of system setup and file management
Advanced Might-green algorithm for automatic measurements and interpretation -
Large internal memory which can be enlarged by USB flash disk & SD card
-
Barcode scanner support
-
LAN/Wi-Fi connections
-
VCG/TVCG/VLP/HRV/Pharmacological Study
-
Supporting DAT、SCP、PDF、XML、HL7、DICOM File format Measurement function and interpretation function
CONFIGURATION
Model
7inch LCD
10.1inch LCD
9 channel
12 channel
15 channel
18 channel
e1200
YES
YES
YES
NO
NO
NO
e1200 plus
YES
YES
YES
NO
NO
NO
YES
YES
YES
YES
YES
NO
e1800
e1800 plus
YES
YES
YES
YES
YES
NO
STANDARD CONFIGURATION
e1200/e1200 plus
Standard Configuration
Part Number
Power cord(European)
07.01.000001
Power adapter
08.01.000001
12-lead Patient Cable (European)
07.03.000014
Adult Chest electrodes (6pcs/set)
08.05.000005
Adult Limb electrodes (4pcs/set)
08.05.000006
ePCECG-12
sampling box
11.1007.0001
USB cable
07.03.100001
Thermal Recorder Paper
(210 mm×295 mm ×100
pages)
08.01.000002
e1800/e1800 plus
Standard Configuration
Part Number
Power Cord (European)
07.01.000001
Power adapter
08.01.000001
18-lead Patient Cable (European)
07.03.000016
Adult Chest electrodes
08.05.000005
Adult Limb electrodes
08.05.000006
ePCECG-18
sampling box
11.1007.0002
USB cable
07.03.100001
Thermal Recorder Paper
(210 mm×295 mm ×100
pages)
08.01.000002
PERFORMANCE SPECIFICATION
-
Anti-electric-shock type : Class I equipment
-
Category AP/APG equipment :
Not category AP/APG equipment
-
Emergency care equipment : Not emergency care equipment
-
Mode of operation: continuous
-
Recorder: Thermal dot-matrix recorder
-
Printing Density :
8 dots per mm / 200 dots per inch (amplitude axes)
40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s) -
Recorder Paper:
Folded thermal paper:
210 mm×295 mm ×100 pages,
215 mm×280mm×100 pages (optional) -
Leads: 9/12/15/18 Leads
-
Gain: 2.5mm/mV、5mm/mV、10mm/mV、
20mm/mV,10/5mm/mV, AGC±5%
-
Speed: 2.5mm/mV、5mm/mV、10mm/mV、20mm/mV,10/5mm/mV, AGC±5%
-
Sampling Frequency: 16KHZ
-
A/D Resolution: 24bits
-
Anti-electric-shock degree: CF type with defibrillation-proo
-
HR Range: 30~300BPM
-
Accuracy: 1BPM
-
Time Constant: ≥3.2s
-
Input Circuit Current: ≤0.01μA
-
Input Voltage Range: ≤±5mVp-p
-
Calibration Voltage: 1mV±2%
-
DC Offset Voltage: ±600mV
-
Input Impedance: ≥50MΩ(10HZ)
-
CMRR: ≥120dB
-
Frequency Response: (0.01~300Hz) +0.4dB / -3.0dB
-
Noise: ≤12.5μV (P-P)
-
AC Filter: 50Hz/60Hz/Off
-
DFT Filter: 0.01HZ/0.05HZ/0.32Hz/0.67Hz
-
Defibrillation-proof recovery time: ≤5s
-
EMG Filter/Lowpass:
25Hz/35Hz/45Hz/75Hz/ 100Hz/150Hz/270Hz/300Hz
-
Pacemaker Detection:
Amplitude: ±750uV~±700mV
Width: 50us~2.0ms
Sampling Frequency: 16000HZ
-
Patient Leakage Current:
NC: ≤10uA(AC)/ ≤10uA(DC)
SFC: ≤50uA(AC)/ ≤50uA(DC)
-
Patient Auxiliary Current:
NC: ≤10uA(AC)/ ≤10uA(DC)